This message is commonly caused by a defective membrane or dried electrolyte. This video tutorial will guide you through the steps to resolve the error message.

Possible Cause Recommended Action
Membrane related (defective or dried up electrolyte) Change the membrane:
1) if it is damaged or missing, has a loose fit, or if there is trapped air or dry electrolyte under the membrane
2) sensor calibration takes almost 14 minutes or cannot be completed within 14 minutes
PCO2 electrode and/ or white PO2 sensing spot contaminated If after having changed the sensor membrane the calibration still takes almost 14 minutes or cannot be completed within 14 minutes contact qualified service personnel to clean the sensor without membrane.
Sentec TC Sensor defective If the problem persists the Sentec TC Sensor is defect. Replace it with a new one.
For more troubleshooting information, see our Service Manual.

 

Related Posts

  • Fehlerbehebung und Verbesserung der Sensorleistung

    LIVE WEBINAR [...]

  • Trainingsmodul

    tCOM+ e-learning modules

    Explore our collection of e-learning modules for using the tCOM+

  • Artikel

    Sensor Problem 12

    Step by step instructions to resolve sensor problem 12 error message